Micro-Analyzer with Optical Detection Micro-Analyzer with Optical Detection

DOI : 10.17577/IJERTCONV3IS25013

Download Full-Text PDF Cite this Publication

Text Only Version

Micro-Analyzer with Optical Detection Micro-Analyzer with Optical Detection

NCRAEEE-2015 Conference Proceedings

Prof.(Dr.) Sakuntala Mahapatra1, Sumant Kumar Mohapatra2+, Sushil Kumar Mahapatra3, Sukant Behera4

Dept. Of ETC, TAT, Bhubaneswar123, Dept of ETC, GIFT, Bhubaneswar4

Abstract:- There is a need to monitor chemical variables online in the environment and in industry. Novel on- line strategies are needed to optimize the measurement and control of these variables. This calls for the development of instrumentation systems capable of in situ measurements. This paper describes the development of a chemical micro-analyzer with an optical detection. The structure was manufactured using low temperature co-red ceramics (LTCC) technology. The optical detection and the ow systems were integrated modularly to micro-analyzer and controlled by wireless communications. The contin- uous ow system was implemented with miniaturized actuators to provide autonomy and portability. The response of the optical detection was evaluated using a dilutions system by multi-commutation. The result was validated with a commercial instrument. The development described and proved in this paper would open up opportunities to new applications with wireless chemical micro-systems networks, a eld with great potential.

Keywords: Micro-analyzer LTCC,Optical detection,Miniaturized ow system, Wireless communication

  1. INTRODUCTION

    Environmental problems as a result of human activities like pol- lution of air, soil, surface water and groundwater have spurred the development of systems for the monitoring, and control of environmental parameters. These systems are also useful for risk assessment and for the evaluation of remediation measures.

    Additionally, online systems have proved to be useful monitor- ing and controlling chemical variables in industry. Nowadays this monitoring is achieved frequently off-line. A sample is taken, trans- ported to a central laboratory to be stored, processed, analyzed until a report is issued and a control decision is made. This takes a lot of time so, when the control action is taken, the control action lags behind the variable of interest. This drives the development of in situ instrumentation systems, with high selectivity and sensitiv- ity, preferably autonomous, capable of being applied to more than one type of sample that allows for simple modication and adap- tation of the analytical method enabling on-line measurements in real time.

    Since the 20th century, analytical systems have reduced steadily the size of the sample [1]. The overall size of the system has been reduced as well thus arriving to the concept of micro-total analy- sis system, µTAS [2]. Nowadays, the automation of these systems includes the micro-uidic optimization and a decrease of opera- tional costs as reagent use is reduced [3]. Micro-systems integrate one or more analytical steps in a small scale including: sampling, sample transport, separation, reaction, measurement, transduc- tion, signal acquisition and processing [4]. Moreover, ow injection systems generally use large actuators like peristaltic pumps and injection valves, resulting in large systems that usually need to be housed in special facilities limiting portability. Several alterna- tives have been reported to optimize uid management [57]. A common approach to micro-systems manufacturing involves sil- icon, glass or polymers requiring costly facilities and procedures. LTCC manufacturing, also known as green tape is advantageous to silicon, glass and polymer fabrication in several ways [8]. LTCC (low temperature co-red ceramics) fabrication is an economi- cal, simple and versatile manufacturing technique. It can produce tridimensional structures for uid management and it can also be a substrate for electronic printed circuits [9,10]. This material and its associated fabrication techniques are used increasingly in the elds of environmental and biomedical monitoring [1114] and also in the development of several types of sensors and actuators [15]. A popular automated analytical method is ow injection analysis, or FIA, as it can produce measurements in real time [16]. To achieve miniaturization, it is necessary to evalu- ate the operational characteristics of selected detection technique, including: sensitivity, small sample volumes (pL, nL, µL) at low con- centrations [17], reliability, short response time, low power and autonomy. Optical methods have a great acceptance in qualita- tive and quantitative determinations. A detection system based on a light emitting diode (LED) and a photodetector offers ver- satility, reliability, accuracy, small size, low cost and low power consumption that can be used in spectrophotometric methods [18]. Robust analyzers have been developed by integrating other devices

    NCRAEEE-2015 Conference Proceedings

    Fig. 1. LTCC manufacturing process.

    such as optical bers [19,20], photomultiplier tubes [21], photo- transistors [22], photodiodes [2325] and reversed biased LEDs [19,2628].

    The need for chemical information in the monitoring and con- trol of chemical processes require real time and on-line response. With the development of wireless communications it is possible to measure in remote locations, far removed from a central laboratory. The integration of wireless communication with analytical instru- ments it is possible to produce devices for in situ measurements [29]. The development of smart sensors called motes has been reported recently. Motes are autonomous nodes with capability of micro-process, signal acquisition, data transmission, battery power management and the communication with others motes [30,31]. A system formed by a network of connected motes linked to a remote server by a gateway is a kind of wireless sensor network (WSN). In the present work, a micro-analyzer fabricated on LTCC (low temperature co-red ceramics) technology featuring optical detection and integrating Bluetooth communication capabilities

    [32] are described. The transmitted data are acquired, processed, stored and displayed with a virtual instrument. The continuous ow system has been built with miniaturized actuators and multi- commutation techniques. The exibility provided by the electronic design allows the micro-analyzer to be operated at different wave- lengths by changing the LED. Also, the sensitivity can be adjusted through the modication of the sampling and commutation times using the virtual instrument. This permits the application of the micro-analyzer to other analytes. The performance of the micro- analyzer was evaluated by determining concentration of KMnO4 . The conguration and control were made by wireless communica- tions.

  2. Experimental

      1. Materials and equipment

        The micro-analyzer was manufactured using LTCC technology featuring DuPont 951PX green tapes (DuPont Microcircuit Mate- rials, NC, USA). The in/out uid connectors on the LTCC structure were made of titanium (MAFSA, S.A de C.V., Torreon, Mexico) and glued using epoxy. The measurement cell was integrated inside the micro-uidic structure and it was covered with a 0.1 mm thick borosilicate glass that allows light beam to pass through the cell. The borosilicate glass allows 90% transmittance whenever working at wavelength between 400 nm and 2 µm [7], and it will not affect the absorbance measurements.

        For the continuous ow system piezoelectric micro-pumps (Takasago Electric, Inc., Nagoya, Japan) were used, dimensions

        25 mm × 25 mm × 4.8 mm. All external connecting tubes were polypropylene with a 1.5 mm internal diameter. The solution were

        injected by multi-commutation [3335] using two-way solenoid micro-valves (161T010, NResearch, NJ, USA).

        The detection system was implemented using an ultrabright LED (Kingbright Co., LA, USA) with a wavelength h = 525 nm. The detector was a TAOS TSL257-LF, light to voltage con- verter (TAOS Inc., Plano, USA) with an integrated transimpedance amplier.

        The control electronic circuit was based on the micro-controller PIC18F4550 (Microchip Technology Inc., Chandler, USA). Surface- mount-device (SMD) electronic components were used to reduce the dimensions of the board. To enable the portability of the system, it was powered with a lithium-ion rechargeable battery VSO-F550H (Amstron PS, Valencia, USA) of 7.2 V2.2 A/h. Wireless communi- cations between the device and control electronics was achieved using Bluetooth modules LM-058 (LM Technologies, London, UK).

        The micro-analyzer was evaluated using dilutions of KMnO4 (SigmaAldrich Co., St. Louis, USA) and bi-distilled water. A 0.05753 g L1 of KMnO4 was prepared as stock solution.

      2. Manufacturing of the micro-uidic structure

        The design of a uidic structure in LTTC follows a layered approach. Every single layer is designed and then machined individually using a CNC mill (LPKF Laser & Electronic, Garbsen Germany). When all layers are stacked and aligned, the desired structure is achieved. The stacked layers are laminated using an

        automated hydraulic press (MAFSA, Torreon, Mex.) with hot plates (100 C) at 750 psi prior to sintering. The laminated LTCC is sintered

        in a programmable furnace (Cole Palmer, Vernon Hills, USA). The general fabrication process for LTCC devices has been described in detail elsewhere [8,14,36]. A simplied diagram of the process is shown in Fig. 1.

        A micro-analyzer may integrate micro-uidic, control and data acquisition electronics or it may follow a modular approach to facil- itate maintenance. The micro-uidic structure is a passive laminar micro-mixer [37] with an integrated detection cell, composed of eight layers: the two top layers where four-liquid inlet/outlet ports are located, four layers to dene the geometry of the channels and

        two bottom layers. Each layer is 0.254 mm (±13 µm). After sinter- ing, the channel dimensions were 0.873 mm (±0.02 mm) wide and

        0.85 mm (±0.013 mm) depth. The dimensions of the detection cell were 3.5 mm × 1.7 mm × 1.72 mm for a 10.5 µl to total volume. The micro-analyzer had a total area of 45 mm × 30 mm and is shown in Fig. 2.

        R.M. Camarillo-Escobedo et al. / Sensorsand Actuators A 199 (2013) 181186

        ISSN: 2278-0181

        NCRAEEE-2015 Conference Proceedings

        Fig. 2. Micro-analyzer with optical detection and wireless communications. (A) Micro-analyzer (45 mm × 30 mm). (B) Miniaturized ow management; 1: Micro-valves; 2: micro-pumps. (C) Internal geometry; 1: inlet/outlet; 2: ow and detection cell (3.5 mm × 1.7 mm × 1.72 mm). (D) Electronic control and wireless modules; 1: intensity

        control; 2: regulators.

      3. Continuous ow system

        Autonomous and portable analytical systems need to realize measurements in real time at the location where the analyte is. The ow system was designed with micro-actuators to improve mixing conditions and minimize reagent use and waste genera- tion. This implementation achieves a smaller, lighter and less power demanding system. Two piezoelectric micro-pumps were cong- ured at 5 Hz of frequency (from 0 to 60 Hz) in order to provide a

        ow rate of 1.5 mL min1 at discharge pressure 5.6 kPa. The micro-

        pumps suck and discharge at the reverse timing. Two micro-valves were congured at 200 ms multi-commutation speed. The injec- tions parameters were set and controlled by a virtual instrument program. In the present work micro-actuators replace bulky pumps and injection valves normally used in ow injection analysis (FIA). This brings our developed system closer to the autonomy and port- ability needed for in situ analysis.

      4. Implementation of the electronic control

        Control data acquisition and signal processing electronics were based on a PIC micro-controller. The system presented here also includes wireless communications between a personal computer (PC) and the micro-controller. The wireless modules use a serial communication and it is converting to Bluetooth for transmitting signals. The serial micro-controller port was used to connect one module and the other module was connected using the PC USB port. Voltage supplies were 5 V for the micro-actuators and 3.3 V for the micro-controller. A start button and a reset button were included in the circuit. The start button runs an activation program prior to normal functioning. The reset button can be pressed at any time to terminate the program and to start over. In each of the inputs of the micro-pumps there were drive circuits that translate the frequency provided by the micro-controller to a signal that drives the micro- pumps to the desired ow rate. The micro-controller also controls

        in PIC-CCS. The micro-controller generates the control frequencies to drive the piezoelectric micro-pumps and sets the times for the opening and closing of the micro-valves in multi-commutation.

      5. Optical detection system

        In spectrophotometry, the radiation emitted or absorbed by a sample occurs in a very short time. The radiation source is a contin- uous beam of light with the relevant wavelength. Any uctuations in the beam intensity can affect measurements. Fluctuations in light intensity in an optical detection system such as the one used in the present work may be due to voltage supply variations. The power supply used in the present work is a high-capacity (2200 mA h) lithium-ion battery followed by a voltage regulator. This ensures a steady voltage that will produce a steady light intensity. In addition, after each measurement, the detector was calibrated using bi- distilled water as reference. As a radiation source an ultrabright LED

        with a h = 525 nm, a diameter of 5 mm with an illumination angle of 20 and transparent plastic bodies was employed. The power

        emitted as light is directly proportional to the current through the LED. The LED emission power is 25 mW and the current required to develop its light intensity is 20 mA. The maximum supply volt- age is 5 V, so the resulting limiting resistance is 32 K. A variable resistor was placed to control the intensity of radiation emitted

        by LED. The photodetector TAOSL-257-LF is a light-to-voltage con- verter. Its output voltage is directly proportional to light intensity (irradiance) on the photodiode. The active area has a round shape with a diameter of 0.75 mm and is located at the center of the lens [38]. When light strikes the photodiode, this current is converted to voltage by a transimpedance amplier integrated in the detector. With this conguration the intensity of incident light is propor- tional to the output voltage of the detector. The detector and the LED were xed in a coupler with two cavities, one where the LED was xed and the other for the photodiode. The distance between these components was 5 mm. This module was placed externally

        the optical detection system including the acquisition of the detec- and perpendicularly to the micro-analyzer cell to achieve maxi-

        tor response. The program for the micro-controller was developed

        mum light measurement. Using this setup the components may

        NCRAEEE-2015 Conference Proceedings

        Fig. 3. Block diagram of experimental set-up.

        be replaced if necessary, without altering the electronic assem- bly.

      6. Procedure

    The micro-analyzer was covered with a black box to isolate the detection system from ambient light avoiding interference from spurious signals. The micro-analyzer was congured from a vir- tual instrument running in the PC. The process begins by activating the wireless communication system to transmit and receive sig- nals via Bluetooth. The communication portwas selected on virtual instrument. Once communication is established micro-actuators are activated rst and then the detection system. Hydrodynamic variables such as ow rate and injection volume of the solutions are adjusted and controlled through a virtual instrument interface by wireless. The dilutions are generated by multi-commutation and transported to the detection cell where the analysis is done obtain- ing an analytical signal that corresponds to the light transmitted and thus to the concentration of the solution. This analytical signal is acquired by an A/D converter in the micro-controller from the sensor and transmitted to the PC via Bluetooth. The virtual instru- mentation program displays the results by a time/voltage graph and stored them in a text le and then processed. The absorbance value was calculated by the Eq. (1) [39]:

    The experiment was made to determine the absorbance of the dilutions obtained by the micro-analyzer. Solutions at differ- ent concentrations were prepared automatically in micro-analyzer from the stock solution of KMnO4 . Measurements were made using a Perkin Elmer Junior III Spectrophotometer 6/8, (Perkin Elmer, Waltham, USA) for comparison purposes.

  3. Results and discussion

    1. Micro-analyzer evaluation

      In the development of miniaturized analytical autonomous sys- tems using ow injection methods, the mixing of solutions may entail the occurrence of a chemical reaction [40,41]. In this work multi-commutation was congured by wireless to prepare dilu- tions from a stock solution and no reaction took place. The solenoid micro-valves switching between potassium permanganate (valve V1) and bi-distilled water (valve V2). Once the dilutions process was validated in the micro-analyzer, the optical system was tested. The intensity of radiation emitted by LED was 4.24 V. It was mea- sured by the photometer TAOSL-257-LF during the course of 5 days. A RSD = 0.002% (n = 20; 98% condence) was obtained. The rela-

      tionship between output voltage and output noise voltage for the voltage regulator used is 10 µV V1 or 10 ppm [42].

      d

      d

      sa sd A = log sr s

      (1)

      The KMnO4 solution has a maximum absorption wavelength of 530 nm, very close to that emitted by the LED (525 nm) that was used. For this evaluation, ow rate for the stock solution and bi-

      where Sa is the analytical signal, Sr is the reference signal measuring bi-distilled water and Sd is the dark signal (signal intensity obtained with the LED turned off).

      The micro-analyzer was evaluated using an experimental set-up shown as a block diagram in Fig. 3.

      distilled water was set at 1.5 mL min1 . Sampling time was set at 8 s to develop a volume of 200 µL according to a ow rate. To pro-

      duce the in-line dilutions automatically, the commutation time was congured between V1 and V2 in a ratio proportional to the dilution required during the sampling time (8 s). The dilutions were made in

      0.8

      0.7

      THEORETICAL

      R.M. Camarillo-Escobedo et al. / Sensorsand Actuators A 199 (2013) 181186

      NCRAEEE-2015 Conference Proceedings

      micro-analyzer can work 55 days. This result suggests the potential

      ConcentraUon KMnO4 (mM) Spectrophotometer

      ConcentraUon KMnO4 (mM) Spectrophotometer

      0.6

      0.5

      0.4

      0.3

      0.2

      0.1

      0

      Microanalyzer vs.Spectrophotometer

      Linear (Microanalyzer vs.Spectrophotometer)

      y = 0.9865x + 0.0055 R² = 0.9993

      of micro-analyzer as a portable system.

      4. Conclusions

      A novel micro-analyzer using optical sensing and wireless trans- mission was developed using low temperature co-red ceramics (LTCC). The resulting system is a step toward the implementation of autonomous analytical systems for in situ determination of envi- ronmental variables and for the control of industrial processes. This micro-analyzer was used to measure KMnO4 using a wavelength of 525 nm as an application example.

      0 0.2 0.4 0.6 0.8

      ConcentraUon KMnO4 (mM) Microanalyzer

      Fig. 4. Comparative response between micro-analyzer and spectrophotometer at

      h = 525 nm.

      a range from 0.238 to 19.6 mg mL1 of KMnO4 . After each measure- ment, the detector was calibrated using bi-distilled water as blank or reference. A linear response was observed. The micro-analyzer equation that describes the linear relationship between concen- tration (KMnO4 ) and absorbance (A) is A = 0.033[KMnO4 ] + 0.0211;

      r2 = 0.9967. For a concentration of 0.734 mM, the RSD obtained was

      1.2% (n = 4; 95% condence). The detection limit was estimated as

      0.139 mM (RSD (%) = 2.16, (n = 4; 95% condence)). The result was compared with measurement made with a spectrophotometer at wavelength 525 nm.

      The absorbance measured by the micro-analyzer was contrasted against measurements made using a Perkin Elmer Junior III Spec- trophotometer 6/8, (Perkin Elmer, Waltham, USA). When both results are plotted against each other, the resulting line has a slope close to unity (0.9993) showing that both methods are equivalent. This is shown in Fig. 4.

      According to the result described above, the micro-analyzer in this work measured KMnO4 concentrations very close to the mea- surement of the spectrophotometer.

      The LTCC micro-analyzer using wireless communications for control and measurements produced results that are equivalent to those produced by a spectrophotometer. The implementation of micro-analyzer is portable and may be used for in situ measure- ments.

    2. Electrical and reagent consumption evaluation

The ow system was designed with miniaturized actuators to improve mixing conditions and to minimize reagent consump- tion as well as waste generation. The automation ow system reduced the reagent consumption, sample consumption and waste generation 25-fold compared with the batch procedure using a spectrophotometer with a 5 mL cuvette. This is an environmentally friendly alternative and could be considered as a step toward green chemistry.

The power consumption has been 120 mA in standby system with micro-pumps, electronic control, wireless communication modules and optical system. Each micro-pump uses 18 mW, its driver circuit uses 240 mW while driving two micro-pumps, for micro-valves the power required is 1 W. In this application it was necessary to operate two micro-pumps and the switching of two micro-valves, with a total of 1.2 W. The power consumption with the operating system was 1.8 W. The dilution procedure was accomplished in 8 s. Therefore, the battery-operated micro- analyzer can achieve 2640 sampling cycles. With a sampling rate of 2 measurements per hour the autonomous ow system of

The main feature of the developed micro-analyzer is the inte- gration of a miniaturized ow system, including micro-pumps and micro-valves for managing uids and a Bluetooth wireless transmission system. It includes a Li-ion power source. The minia- turization of actuators (valves, pumps) and other components (sensor, electronics, power supply) in this micro-analyzer together with reduced power consumption and small reagent volumes, result in an autonomous portable system.

The result of this application could lead to advances in the devel- opment of wireless chemical micro-systems networks.

References

  1. M. Varcarcel, M.S. Cardenas, Automatización y miniaturización en Química Analítica, Springer, Córdoba, 2001.

  2. A. Manz, N. Graber, H.M. Widner, Miniaturized total chemical analysis systems: a novel concept for chemical sensing, Sensor and Actuator B-Chemical (1990) 244248.

  3. D.R. Reyes, D. Iossidis, P. Aroux, A. Manz, Micrototal analysis system. Introduc- tion, theory and technology, Analytical Chemistry 7 (2002) 26232636.

  4. S. Alegret, Integration, a new paradigm in analytical chemistry, in: D. Integrated, Barcelo (Eds.), Analytical Systems, Elsevier, Amsterdam, 2003, pp. 133.

  5. E.A. Zagatto, B.F. Reis, C.C.Oliveira, R.P. Sartini, M.A.Z. Arruda, Evolution of the communication concept associated with the development of ow analysis, Analytica Chimica Acta 400 (1999) 249256.

  6. R. Camarillo, F. Valdés, J. Alonso, Desarrollo de un Sistema Hidrodinamico para Sistemas de Análisis en Flujo Miniaturizados, Revista Mexicana de Ingeniería Química 11 (2012) 299307.

  7. M.M. Baeza, Nuevas estrategias para la gestión de los uidos en sistemas autom- atizados de análisis, Tesis de Doctorado en Quimica, Universidad Autónoma de Barcelona, Espan a, 2004.

  8. M. Gongora, P. Espinoza, L. Sola, J. Santiago, Overview of low temperature cored ceramics tape technology for meso-system technology (MsST), Sensor and Actuator, A-Physical 89 (2002) 222241.

  9. J. Li, G.K. Ananthasuresh, Three-dimensional low-temperature co-red ceramic shelles for miniature system application, Journal of Micromechanics and Micro- engineering 12 (2002) 198203.

  10. Y. Imanaka, Multilayered Low Temperature Cored Ceramics (LTCC) Technol- ogy, Springer, Boston, USA, 2005.

  11. N. Iban ez, Miniaturización de Analizadores Químicos mediante la tecnología LTCC, Tesis de Doctorado en Química, Universidad Autónoma de Barcelona, Espan a, 2007.

  12. N. Iban ez, R.D.M. Goncalves, Z.M. da Rocha, R.M. Gongora, A.C. Seabra, J. Alonso, LTCC meso-analylical system for chloride ion determination in drink water, Sensor and Actuator B. Chemical 118 (12) (2006) 67.

  13. C. Martínez, N. Iban ez, F. Valdes, J. Alonso, LTCC Microow analizer with mono- litic integration of thermal control, Sensor and Actuator A: Physical (2007) 6370.

  14. N. Iban ez, J. Alonso, C. Martínez, F. Valdes, Green tape ceramics new techno- logical approach for integrating electronic and uidics in microsystems, TrAc Trends in Analytical Chemistry (2008) 2433.

  15. K.A. Peterson, K.D. Patel, C.K. Ho, S.B. Rohde, C.D. Nordquist, C.A. Walker, B.D. Wroblewski, M. Okandan, Novel microsystem application with new techniques in low temperature co-red ceramics, Intermational Journal Applied Ceramics Technology 2 (5) (2005) 345363.

  16. K.N. Andrew, N.J. Blundell, D. Price, P.J. Worsfold, Flow injection techniques for water monitoring, Analytical Chemistry A (1994) 916922.

  17. C.B. Boring, P.K. Dasgupta, An affordable high-performance optical absorbance detector for capillary systems, Analytical Chimica Acta 342 (1997) 123132.

  18. D. Diamond, Internet-scale sensing, Analytical Chemistry 76 (2004) 279A286A.

  19. K. Toda, K.I. Yoshioka, S.I. Ohira, J. Li, P.K. Dasgupta, Trace gas measurement with an integrated porous tube collector/long-path absorbance detector, Analytical Chemistry 75 (2003) 40504056.

  20. G. Samanta, C.B. Boring, P.K. Dasgupta, Continuous automated measurement of hexavalent chromium in airboorne particulate matter, Analytical Chemistry 73 (2001) 20342040.

  21. G.E. Collins, Q. Lu, N. Pereira, P. Wu, Long pathlength, three-dimensional absorbance microchip, Talanta 72 (2007) 301304.

  22. M.A. Fores, B.F. Reis, A downsized ow set up based on multicommuta- tion for the sequential photometric determination of iron(II)/iron(III) and nitrite/nitrate in surface water, Talanta 68 (2005) 422428.

  23. A. Rainelli, R. Stratz, K. Schweizer, P.C. Hauser, Miniaturized ow-injection analysis manifold created by micromilling, Talanta 61 (2003) 659665.

  24. N. Gros, Spectrometer with microreaction chamber and tri-colour light emit- ting diode as a light source, Talanta 62 (2004) 143150.

  25. A. Fonceca, J. Raimundo, M. Ivo, A multichannel photometer based on an array of light emitting diode for use in multivariate calibration, Analytical Chimica Acta 522 (2004) 223229.

  26. M. OToole, K.T. Lau, D. Diamond, Photometric detection in ow analysis sys- tems using integrated PEDDs, Talanta 66 (2005) 13401344.

  27. K.T. Lau, S. Baldwin, M. OToole, R. Shepherd, W.J. Yerzunis, S. Izuo, S. Ueyama, D. Diamond, A low-cost optical sensing device based on paired emitter-detector light emitting diodes, Analytica Chimica Acta 557 (2006) 111116.

  28. K.T. Lau, R. Shepherd, D. Diamond, Solid state ph sensor based on light emitting diodes (LED) as detector plataform, Sensors 6 (2006) 848859.

  29. D. Diamond, K.T. Lau, S. Brady, J. Cleary, Integration of analytical measurements and wireless communications current issues and futures strategies, Talanta 75 (2008) 606612.

  30. B. Warneke, S.J. Kristofer, D. Smart, Comunication with a cubic-multimeter computer, Classical Papers on Computational Logic 1 (2001) 372383.

  31. K. Yao, Sensor networking: concepts, applications and challenge, Acta Auto- matica Sinica 32 (6) (2006), 939845.

  32. Volume 2: Core System Package [Controle Volume]. Especicación, v2.0 Blue- tooth SIG, Noviembre 2004.

  33. F. Rocha, B. Reis, E. Zagatto, J. Lima, R. Lapa, J. Santos, Multicommutation in ow analysis: concepts, applications and trends, Analytica Chimica Acta 468 (2002) 119131.

  34. E. Rodena, A. Morales, M. De la Guardia, Determination of phenols in waters using micro-pumped multicommutation and spectrophotometric detection. An automated alternative to the satndard procedure, Analytical and Bioana- lytical Chemistry 383 (1) (2005) 138144.

  35. F. Rocha, E. Rodena, A. Rubio, M. De la Guardia, A clean method for ow injection spectrophotometric determination of cyclamate in table sweeteners, Analytica Chimica Acta 547 (2005) 204208.

  36. R. Camarillo, F. Valdés, J. Camarillo, La Tecnología LTCC aplicada al Desarrollo de Sistemas de Microuidica, in: 2010 Presentación A3 121. XVI Congreso Inter- nacional anual de la Sociedad Mexicana de Ingeniería Mecánica SOMIM, 2224 Septiembre, Monterrey, NL, 2010.

  37. N.T. Nguyen, S. Wereley, Fundamentals and Aplicationsof Microuidics, Artech House, Boston, 2006.

  38. Datasheet TSL257. High Sensitiviti Light-to-Voltage Converter, TAOS.

  39. A. Fonseca, I. Raimundo, A multichannel photometer based on an array of light emitting diodes for use in multivariate calibration, Analytica Chimica Acta 522 (2004) 223229.

  40. J. Ruzicka, J.W. Stewart, Flow injection analysis, Analytica Chimica Acta 79 (1975) 7991.

    NCRAEEE-2015 Conference Proceedings

  41. M. Varcarcel, M.D. Luque de Castro, Analisis por inyeccion en ujo., Imprenta Sn. Pablo, Murcia, Córdoba, 1984.

  42. Fairchild Semiconductor Corporation, Datasheet KA78XX/KA78XXA 3 terminal 1A positive voltage regulator, 2001.

Leave a Reply