Is the Depth Worth It? A Cost-Aware Comparative Study of ResNet18 and ResNet50 for Brain Tumor MRI Classification and Clinical Deployment

DOI : 10.5281/zenodo.21755806
Download Full-Text PDF Cite this Publication

Text Only Version

Is the Depth Worth It? A Cost-Aware Comparative Study of ResNet18 and ResNet50 for Brain Tumor MRI Classification and Clinical Deployment

Asutosh Dash, Varsha Rajendran, Ritesh Ranjan Panda

SCOPE

VIT-AP University Vijayawada, Andhra Pradesh

Abstract – Brain tumors are a health concern worldwide, and timely diagnosis increases the probability of survival for pa- tients. In the present research, two deep learning architectures, ResNet18 and ResNet50, are compared and evaluated in terms of brain tumor recognition, and the question of whether the higher training and inference cost of the deeper model is justified is studied. Preprocessing methods are applied to clarify images and improve model performance before training. Accuracy, precision, recall, confusion matrix, Grad-CAM interpretations, and com-putational cost metricsincluding parameter count, model size, FLOPs, training time, and inference timeare considered in the evaluation. The results show a modest, statistically ambiguous performance gain for ResNet50 that is concentrated in a single class (meningioma), while ResNet18 matches or exceeds ResNet50 on the remaining classes at substantially lower computational cost. Trade-offs are discussed along with class- and deployment-specific recommendations.

Index Terms – Brain MRI Classification, Deep Learning, ResNet18, ResNet50, Transfer Learning, Grad-CAM, Brain Tu- mor Detection, Computational Cost, Clinical Deployment, Com- parative Study

  1. INTRODUCTION

    imaging [1]. There are two commonly used variants, ResNet18 and ResNet50; ResNet18 is characterized by faster and less resource- intensive computation, while ResNet50 has greater capacity for feature extraction but requires more computational resources [3]. The choice of the deeper architecture for deploy-ment in a clinical setting is not determined by accuracy alone. Hospitals and diagnostic centers, especially those operating in resource-limited environments, must weigh prediction perfor-mance against inference time, memory consumption, hardware cost, and energy consumption. A model that provides only a marginal improvement in accuracy but requires substantially more resources for inference is not always the best choice for a clinical workflow. Although both models are widely used across numerous studies, a direct, cost-aware comparison under identical conditions for brain tumor classification is rather rare [2], [4]. This lack of cost-aware comparison makes it necessary not only to determine which model performs better, but to determine whether the performance gain of the deeper architecture is worth its additional cost.

    1. Background

      Brain tumors are a critical issue for medicine, and timely MRI scan- based detection is significant for improving patient outcomes. Manual analysis of MRI scans may be erroneous, and therefore the use of deep learning methods for automatic classification is widely spread. Residual Neural Networks (ResNets) are preferred due to their ability to resolve training problems and extract deeper features [1]. Recent studies have shown that ResNet50 combined with an explainable approach like Grad-CAM demonstrates superior performance and inter-pretability for brain tumor detection [2], [3].

    2. Motivation

      Brain tumor detection is a challenging task because review-ing MRI scans can be slow and error-prone. Deep learning models such as Residual Networks have shown excellent results in computer vision tasks and are widely used in medical

    3. Problem Statement

      Brain tumors are a serious health threat, and timely detec-tion is critical for improving patients probability of survival. Even though MRI scans provide detailed images, manual analysis can be slow and error-prone. Deep learning models like ResNet18 and ResNet50 have proven their applicability for automatic brain tumor detection; however, comparative studies under identical experimental conditions are scarce, and only a few studies compare the models while also quan-titatively assessing their computational cost. Evaluating the balance between accuracy, computational performance, and interpretabilityand, critically, whether that balance favors a lightweight or a deeper model in practiceis necessary to identify which architecture is more suitable for real- world clinical applications [1][3].

      1. Objectives

        In the present research, ResNet18 and ResNet50 are de- veloped and evaluated using a publicly available brain MRI dataset that includes images of glioma, meningioma, pituitary tumor, and normal brain. The performance of the models is compared using metrics such as accuracy, precision, recall, and F1-score. In addition, learning curves are examined to under- stand the dynamics of model improvement during training. To make the models more interpretable, the attention behavior of the networks on MRI scans is demonstrated using Grad-CAM. Beyond standard metrics, the study quantitatively evaluates the computational cost of each architecture: parameter count, model size, FLOPs, training time, and inference time.

      2. Contributions

    This paper provides a structured comparison of ResNet18 and ResNet50 for brain MRI classification that explicitly ac- counts for computational cost. The study includes an extensive experimental evaluation using accuracy, confusion matrices, and other classification measures for each tumor type. In addi- tion, the value of an interpretable model is highlighted through Grad-CAM visualizations, revealing the most important MRI regions considered by each model. The paper also offers insight into the balance between cost and prediction accuracy, and questions the significance of the accuracy improvement of ResNet50 over ResNet18.

  2. RELATED WORK

    1. Traditional Machine Learning in Brain MRI Classification

      Early attempts to classify brain MRI relied mainly on handcrafted features combined with classical machine learn- ing algorithms. SVM, RF, k-NN, and other classifiers were used to classify brain tumors based on MRI texture, shape, and intensity. Although these methods showed satisfactory results, their major disadvantage was an inability to capture the multilayered characteristics of medical images, leading to inferior performance compared to modern deep learning approaches [5].

    2. Deep Learning in Brain MRI Classification

      CNNs marked a revolution in medical image analysis since these models can learn important features automatically, with- out manual feature engineering [6], [7]. A number of studies have demonstrated the superiority of CNNs over traditional machine learning approaches for the detection and classifi- cation of brain tumors on MRI [4], [8]. Nonetheless, the performance of early CNNs was restricted by their simple architecture and limited ability to classify diverse tumor types and adapt to varying MRI quality [5].

    3. Residual Networks (ResNets) in Medical Imaging

      ResNet, or Residual Network, was invented to solve the vanishing-gradient problem and enable the training of much deeper models [1]. ResNet18 and ResNet50 are both com- monly used for brain MRI classification. ResNet18 is highly efficient and has a reduced computational cost, making it

      appropriate for live applications and resource-constrained set- tings [5]. ResNet50, owing to its deeper architecture, usually achieves higher accuracy due to improved feature extraction, thoug at the cost of higher memory usage and longer training time [2]. Most research concentrates on either ResNet18 or ResNet50 in isolation, providing little direct comparison of the two under identical conditionsleaving a gap in understand- ing the trade-offs among network depth, prediction accuracy, and performance speed [4], [8].

    4. Explainable AI in Medical Imaging

      Explainable AI has gained increasing importance in medical imaging, since high accuracy alone is not sufficient for clinical applications [9]. The Grad-CAM (Gradient-weighted Class Activation Mapping) technique allows visualization of the MRI regions that contribute most to a models prediction [10]. Research has shown that Grad-CAM can help verify whether a model is focusing on tumor-related regions [2].

    5. Research Gap

    Despite previous attempts to use ResNets for brain tumor classification, most existing papers address the problem by simply comparing the accuracy, precision, recall, or F1-score of two network architectures [2], [3]. If one network per- forms marginally better on any of these metrics, the deeper architecture is typically considered superior, without asking whether the difference is truly significant and reliable, and without weighing that improvement against the additional computational complexity and training/inference cost it incurs. In practice, however, these networks may be deployed in resource-constrained hospitals, point-of-care devices, or high- throughput screening pipelines, where computational cost is a crucial deployment criterion. Comprehensive comparative analyses that jointly report predictive performance and com- putational cost under uniform experimental conditions remain limited [4], [8], and the statistical significance of small ac- curacy differences between architectures is rarely examined. This study takes a step forward by performing a cost-sensitive comparative analysis of ResNet18 and ResNet50: alongside reporting the performance metrics of the two architectures, this paper investigates whether the additional computation required by ResNet50 is justified by its additional performance, using quantitative data on performance metrics, computational cost, and Grad-CAM visualizations.

  3. METHODOLOGY

    1. Proposed Workflow

      The overall workflow followed in this study for comparing ResNet18 and ResNet50 on brain MRI classification can be divided into six main stages, as illustrated in Fig. 1.

      1. Data Preprocessing: The collection of brain MRI im- ages comprised four categories: glioma, meningioma, pituitary tumor, and non-tumor images. The collected dataset was split into portions used to train the model and to evaluate it. To conform to the input size expected by the ResNet architectures,

        Fig. 1. Schematic representation of the end-to-end experimental pipeline consisting of six stages: (1) Dataset preparation, (2) Model selection, (3) Transfer learning setup, (4) Training phase, (5) Evaluation phase, and (6) Model explainability using Grad-CAM. Both architectures are passed through exactly the same pipeline.

        all images were resized to 224 × 224 pixels. Several data aug- mentation techniques were used to improve the generalization power of the training set: (1) random horizontal flipping, (2) random rotations, and (3) normalization using the mean and standard deviation values of the ImageNet dataset.

      2. Model Selection: Two pre-trained CNN architectures were employed:

        • ResNet18: A shallower network containing 18 layers.

        • ResNet50: A deeper network containing 50 layers. Both models were initialized with ImageNet pre-trained weights. Modifications were introduced to the last FC layer so that the output size matched the number of target classes (4).

      3. Transfer Learning Setup: The primary convolutional

        structure in both models was held constant to preserve the fundamental feature representations already acquired. Adjust- ments were restricted to the ultimate classification component, using the set of brain MRI images for this purpose. The Adam optimization algorithm was chosen, configured with a learning rate of 0.001. The objective function used for minimization was the Cross-Entropy measure of error.

      4. Training Phase: Training was carried out over 5 epochs. In each epoch, the forward pass was performed for training batches, the loss was computed and backpropagated, and model parameters were updated. The performance of the models was tracked after each epoch in terms of accuracy and loss values.

      5. Evaluation Phase: After training, both models were evaluated on the validation dataset. The following metrics were computed: accuracy (fraction of correct predictions), precision of positive predictions, recall of positive predictions, and the harmonic mean of precision and recall. Confusion matrices were plotted to illustrate the models mistakes, and graphs illustrating the dynamics of model performance were produced.

      6. Explainability via Grad-CAM: To make the models more understandable, the Grad-CAM methodology (Class Ac- tivation Mapping with Gradient weighting) was used. Grad- CAM visualizations showed which areas of the MRI images the networks attended to when performing classification, thus verifying whether the networks were genuinely focusing on tumor regions.

    2. Preprocessing

      Preprocessing of MRI scans is crucial for classification with deep learning. To ensure consistency and enhance the generalizability of the models, the steps shown in Fig. 2 were carried out.

      Fig. 2. Preprocessing pipeline applied equally to both models: (a) scaling to 224 × 224 pixels, (b) normalization based on ImageNet channel means and standard deviations, and (c) data augmentation (horizontal flips, rotations, and color normalization) applied only to the training subset.

      1. Resizing Images: The size of each MRI image was altered to ensure a resolution of 224 × 224 pixels. This resizing ensures that the images meet the size expected by the ResNet18 and ResNet50 networks, since these architectures are built around the ImageNet standard size.

      2. Normalization: Pixel intensity values in the MRI scans were standardized using the per-channel mean and standard deviation derived from the ImageNet collection:

        • Mean: [0.485, 0.456, 0.406]

        • Standard Deviation: [0.229, 0.224, 0.225]

          This standardization guarantees that the statistical properties of the incoming MRI data match those expected by the models already trained on a large dataset, making knowledge transfer efficient.

      3. Data Augmentation: To lessen the risk of over-

        specialization and improve the resilience of the networks, data enhancement methods were used exclusively on the training data:

        • Random Horizontal Flip (p = 0.5): Reflects the image across the vertical center line with 50% probability, generating additional training diversity.

        • Random Rotation (±10°): Slight random rotations help

          the network learn rotational invariance.

        • Color Normalization: Keeps input channel statistics consistent.

          These preprocessing techniques simulate the effects that can occur in MRI images due to variations in patient positioning, scanning method, and scanner orientation.

    3. Dataset Setup

      The data used in this study consists of brain magnetic resonance images categorized into four classes: glioma, menin-

      gioma, pituitary tumor, and normal brain tissue. Te data is stored in .jpg format and further divided into training, validation, and test sets.

      • Training Subset: Used for fitting the neural network models.

      • Validation Subset: Used for adjusting model config- uration settings and assessing performance consistency throughout the learning phase.

      • Test Subset: Reserved for a single, final assessment of the fully trained networks.

      The class breakdown of the data is presented in Table I.

      TABLE I

      Class-wise Distribution of the Brain MRI Dataset Across Training, Validation, and Test Partitions

      Class

      Train

      Validation

      Test

      Glioma

      1134

      243

      244

      Meningioma

      1151

      247

      247

      No Tumor

      1400

      300

      300

      Pituitary

      1229

      264

      264

      Total

      4914

      1054

      1055

    4. Model Architectures

      1. ResNet18 Architecture:

        • Depth: 18 layers deep.

        • Basic Block: The foundational module consists of a pair of convolution operations integrated with batch normal- ization and ReLU non-linearity, along with a shortcut (skip) connection.

        • Structure:

          • Initial convolution: 7 × 7 Conv + MaxPool

          • 4 residual stages with 2 blocks each

          • Global Average Pooling

          • Fully Connected (FC) layer for classification

      2. ResNet50 Architecture:

        • Depth: 50 layers deep.

        • Bottleneck Block: The key unit is composed of three consecutive convolution operations (1 × 1 followed by a 3 × 3 and another 1 × 1) rather than two, which lowers processing demands while enabling a deeper architecture.

        • Structure:

          • Initial convolution: 7 × 7 Conv + MaxPool

          • 4 residual stages with 3, 4, 6, and 3 blocks respec- tively

          • Global Average Pooling

          • Fully Connected (FC) layer for classification

  4. RESULTS AND DISCUSSION

    1. Classification Report Comparison

      Table II presents the overall performance metrics for both models on the held-out validation set.

      Interpretation: The ResNet50 model shows only marginally better performance than ResNet18about

      0.2 percentage points in overall accuracy (89.09% vs.

      TABLE II

      Overall Classification Performance of ResNet18 and

      ResNet50 on the Validation Set

      Model

      Accuracy

      Precision

      Recall

      F1-Score

      ResNet18

      88.90%

      0.8869

      0.8890

      0.8875

      ResNet50

      89.09%

      0.8990

      0.8909

      0.8928

      88.90%). This is consistent with the architectural rationale: because the bottleneck blocks of ResNet50 allow the stacking of more layers of non-linear transformations at a comparable cost per layer, it is able to learn more discriminative, hierarchical features of MRI texture than ResNet18s basic- block design [1]. However, the gap is very small, and Section IV-F discusses whether this difference is statistically significant or the result of run-to-run variability.

    2. Class-wise Comparison

      Table III presents the F1-scores for each of the four classes.

      TABLE III

      Per-Class F1-Score Comparison Between ResNet18 and

      ResNet50

      Class

      ResNet18 (F1)

      ResNet50 (F1)

      Glioma

      0.894

      0.885

      Meningioma

      0.769

      0.810

      No Tumor

      0.941

      0.942

      Pituitary

      0.931

      0.922

      Interpretation: ResNet50s advantage is not uniform across classes. It clearly outperforms ResNet18 on meningioma (0.810 vs. 0.769 F1)the class most prone to confusion with gliomaconsistent with the idea that meningiomas heteroge- neous appearance benefits from ResNet50s deeper feature hi- erarchy. However, ResNet18 slightly outperforms ResNet50 on both glioma (0.894 vs. 0.885) and pituitary (0.931 vs. 0.922), and the two models are essentially tied on No Tumor (0.941 vs. 0.942). This indicates that ResNet50s added depth helps specifically with the hardest class to separate (meningioma), but does not uniformly improve performance elsewhere, and in two of four classes a shallower network generalized marginally better.

    3. Confusion Matrix Analysis

      The confusion matrices for ResNet18 and ResNet50 offer a more detailed understanding of efficacy at the individual class level, as shown in Fig. 3.

      1. Glioma: ResNet18 correctly classified 216 out of 243 glioma cases, with 26 misclassified as meningioma and 1 as pituitary. ResNet50 correctly classified 204 out of 243 glioma cases, with 36 misclassified as meningioma and 3 as pituitary. ResNet18 shows both higher recall and fewer total confusions on this class than ResNet50the deeper network did not improve glioma discrimination in this run.

        Fig. 3. Validation-set confusion matrices for (left) ResNet18 and (right) ResNet50, with rows denoting ground-truth class and columns denoting predicted class. Diagonal entries indicate correct classifications.

      2. Meningioma: ResNet18 correctly classified 183 out of 247 meningioma cases, with confusion toward glioma (22), No Tumor (22), and pituitary (20). ResNet50 correctly classified

        218 out of 247 meningioma cases, with confusion toward glioma (8), No Tumor (13), and pituitary (8). ResNet50 shows a substantial recall improvement on meningiomathe class where it misclassifies noticeably fewer cases toward every other category, including a marked drop in confusion with glioma (8 vs. 22).

      3. No Tumor: ResNet18 correctly classified 288 out of 300 healthy cases, with 2 misclassified as glioma, 8 as menin- gioma, and 2 as pituitary. ResNet50 correctly classified 282 out of 300 healthy cases, with 3 misclassified as glioma and 15 as meningioma. Both models perform very well on this class, with ResNet18 marginally ahead and producing fewer false negatives.

      4. Pituitary Tumor: ResNet18 correctly classified 250 out of 264 pituitary cases, with 12 misclassified as meningioma and 2 as No Tumor. ResNet50 correctly classified 235 out of 264 pituitary cases, with 3 misclassified as glioma, 22 as meningioma, and 4 as No Tumor. ResNet18 shows a clear advantage on this class, with roughly half as many pituitary cases confused with meningioma compared to ResNet50.

      5. Summary: The confusion matrices show a more mixed picture than a simple the deeper model is better narra- tive. ResNet50s only clear class-level advantage is menin- gioma, where it reduces confusion with every other class, most notably glioma. For glioma, No Tumor, and pituitary, ResNet18 achieves equal or higher recall and produces fewer total misclassifications. This reinfoces the papers central argument: ResNet50s added depth pays off specifically for the hardest class to separate, but does not translate into a uniform improvement across the task, and in three of four classes the shallower network performed as well or better in this run.

    4. Accuracy and Loss Curves

      The performance curves of ResNet50 and ResNet18 show that ResNet50 achieves marginally higher final accuracy, with both models converging to a similar level by epoch 5, as illustrated in Fig. 4.

      ResNet18: Training accuracy improves steadily from 75.46% (epoch 1) to 86.61% (epoch 5), while validation accu- racy improves monotonically from 82.07% to 88.90% over the

      Fig. 4. Training (blue) and validation (red) accuracy curves over 5 epochs for (left) ResNet18 and (right) ResNet50.

      same period. The gap between training and validation accuracy stays small and closes further by the final epoch, indicating good, stable generalization with no signs of overfitting.

      ResNet50: Training accuracy improves from 74.99% (epoch 1) to 87.40% (epoch 5), while validation accuracy moves from 83.78% to 89.09%but non-monotonically: validation accuracy actually peaked at 87.19% at epoch 3, then dropped to 84.91% at epoch 4 (with validation loss simultaneously rising from 0.3577 to 0.4029), before recovering to its final value of 89.09% at epoch 5. This mid-training instability visible as a visible dip-and-recover pattern in both the accuracy and loss curvesindicates ResNet50s validation performance was less consistent across epochs than ResNet18s smooth, monotonic trajectory in this run.

      In summary, ResNet18 converges smoothly and predictably, while ResNet50 achieves a marginally higher final accuracy but only after a pronounced epoch-4 dip that ResNet18 does not exhibit. As discussed in Section IV-E, ResNet50s per- epoch cost is also substantially higher, since each ResNet50 epoch took roughly 2.5× as long as a ResNet18 epoch in this training runmeaning the deeper model was both less stable during training and considerably more expensive to train.

      Fig. 5. Training (blue) and validation (red) loss curves over 5 epochs for (left) ResNet18 and (right) ResNet50.

    5. Computational Cost Analysis

      The accuracy and F1-score of a model do not, by them- selves, imply deployment readiness. Table IV gives the compu- tational characteristics of both models, based on architecture- level measurements and on the authors own CPU training run.

      *Parameter counts, model size, FLOPs, and single-image (batch=1) infer-

      ence latency were measured directly using PyTorch 2.13 and torchvision (ImageNet-standard ResNet18/ResNet50 definitions with the final FC layer replaced by a 4-class head, matching Section III-D) on a single-core Intel

      TABLE IV

      Computational Cost Comparison Between ResNet18 and

      ResNet50

      Metric

      ResNet18

      ResNet50

      Total parameters

      11,178,564

      23,516,228

      Trainable params (FC head, 4

      2,052

      8,196

      classes)

      Model size on disk (FP32)

      42.72 MB

      90.01 MB

      FLOPs (per 224×224 image)

      3.65 GFLOPs

      8.26 GFLOPs

      Inference latency (batch=1,

      47.6 ms

      102.6 ms

      CPU*)

      Throughput (batch=32, au-

      44.0 img/s

      20.1 img/s

      thors training CPU)

      Training time (per epoch, au-

      153.7 s

      377.6 s

      thors training CPU)

      Total training time (5 epochs,

      768.7 s

      1888.1 s

      authors training CPU)

      Xeon CPU @ 2.10 GHz with CUDA unavailable; batch=1 latency is averaged over 30 forward passes following a 25 pass warm-up. These batch=1 figures are hardware-invariant in relative terms, but absolute latency will differ on other hardware.

      Training time, total training time, and throughput (batch=32) were measured directly on the authors own CPU-only training run (Windows, CPU-only PyTorch), as recorded in the training console log, rather than on the single- core reference CPU used for the batch=1 latency figures above; these two measurement sources are not directly comparable to each other.

      The number of parameters validates the common architecture design [1]: ResNet50 contains almost 2.1× the number of parameters of ResNet18 due to its bottleneck structure and increased stage depths. This translates to a 2.1× larger model size and a 2.3× increase in FLOPs per image. On the authors training hardware, ResNet50 took roughly 2.5× the training time per epoch of ResNet18 (377.6 s vs. 153.7 s) and achieved less than half of ResNet18s batch-32 throughput (20.1 vs. 44.0 images/sec). Costbenefit interpretation: ResNet50 costs roughly twice as much in parameters, model size, and FLOPs, and required approximately 2.5× the training time and less than half the training throughput of ResNet18, for an overall accuracy improvement of only about 0.2 percentage points in the single-run comparison (Table II), or roughly 1.1 per- centage points in the more reliable multi-seed study (Sec- tion IV-F). Given how small this accuracy gain is relative to the computational overhead, ResNet18 appears to be the more cost-effective choice for most deployment scenarios in this study, particularly latency- or resource-constrained settings such as point-of-care screening. ResNet50s primary advantage is class-specific: it substantially reduces meningioma misclas- sification (Section IV-E class-wise results above), which may justify its cost in settings where missing meningioma cases is a specific clinical prioritybut this is a narrower and more conditional justification than a blanket ResNet50 is better conclusion.

    6. Statistical Significance of the Observed Accuracy Differ- ence

      The single-run comparison in Table II shows only a 0.2 per- centage point accuracy difference between ResNet50 (89.09%)

      and ResNet18 (88.90%), which on its own could easily be at- tributable to random initialization, augmentation randomness, or batch ordering rather than genuine architectural superiority. To properly test this, we repeated training for both models across 5 random seeds (4246) and evaluated the resulting distributions.

      Important caveat: the multi-seed runs used a training con- figuration distinct from the main 5-epoch comparison above absolute accuracies across seeds (86.187.8%) are noticeably lower than the 88.989.1% obtained in the single main run reported in Table II, likely reflecting a shorter or otherwise different training schedule used for the seed sweep. The multi- seed results below should therefore be read as evidence about the direction and reliability of the ResNet18-vs-ResNet50 gap, not as a direct replacement for the absolute accuracy figures in Table II.

      Mean ± standard deviation across 5 seeds:

      • ResNet18: accuracy = 0.8612 ± 0.0061, F1 = 0.8602 ±

        0.0047

      • ResNet50: accuracy = 0.8726 ± 0.0087, F1 = 0.8707 ±

        0.0119

        Across the 5 seeds, ResNet50 outperformed ResNet18 in every single seed (mean difference = 0.0114, i.e., ResNet50 higher by about 1.1 percentage points on average), which is a substantially larger and more consistent gap than the 0.2-point difference seen in the single main run.

        Paired significance tests on per-seed accuracy:

      • Paired t-test: t = 3.109, p = 0.036 (significant at = 0.05).

      • Wilcoxon sined-rank test: p = 0.0625 (not significant at = 0.05, though close to the threshold; with only 5 paired samples the Wilcoxon test has limited power to detect an effect).

        The two tests disagree at the conventional 0.05 threshold: the parametric paired t-test indicates a statistically significant difference, while the non-parametric Wilcoxon testmore ap- propriate for small samples with uncertain normality, but also lower-powered with only 5 pairsdoes not reach significance. Given this disagreement and the very small sample size (n = 5 seeds), we regard the evidence for a consistent ResNet50 advantage as suggestive but not conclusive.

        McNemars test on paired predictions (single seed, same test set): Comparing per-image predictions from the two models on identical test samples, 65 images were classified correctly by ResNet18 but incorrectly by ResNet50, while 73 images were classified correctly by ResNet50 but incorrectly by ResNet18 (2 = 0.355, p = 0.551)not significant. This test operates at the level of individual predictions rather than aggregate accuracy across seeds, and its non-significant result indicates that, for this single seeds test-set predictions, the

        two models disagree on a comparable number of cases in each direction, without a clear directional advantage.

        Overall interpretation: the evidence for ResNet50s supe- riority is mixed rather than clear-cut. The seed-level accuracy comparison (paired t-test) suggests a real, if modest, advantage

        for ResNet50, but this is not confirmed by the non-parametric Wilcoxon test or by the prediction-level McNemars test. Com- bined with the single-run result in Table II showing only a 0.2- point gap, we conclude that ResNet50s accuracy advantage over ResNet18, while directionally consistent across seeds, is small in absolute terms and not robustly confirmed across all statistical tests. Any deployment recommendation in this paper is therefore based on the combined weight of this evidence together with the computational cost analysis (Section IV-E), not on a definitive statistical proof of ResNet50s superiority.

    7. Explainability with Grad-CAM

    Even though highly advanced neural networks such as ResNet18 and ResNet50 are extremely effective at detecting brain tumors, the opacity of their internal decision-making normally limits their acceptance in medicine. Besides high accuracy, interpretability of the AI system is also necessary for medical professionals [9].

    One of the most frequently used methods for interpret- ing Convolutional Neural Network decisions is Grad-CAM (Gradient-based Class Activation Mapping). Using the gradi- ent of the predicted class with respect to the final convolutional feature maps, Grad-CAM creates overlays showing the parts of the image most important to the networks final decision [10]. There is ample evidence that Grad-CAM can be applied in practice to interpret brain tumor MRIs. For example, Sahli et al. state that visual explanations make models more transparent and allow radiologists to check whether the AIs decision was correct [3]. Transparency is extremely important in medicine, since AI should serve only as a decision-making aid and not

    as a substitute for a physician.

    In the current research, Grad-CAM is applied to the classi- fication results of both ResNet18 and ResNet50 to ensure the clinical relevance of their decisions. Example visualizations are presented in Fig. 6.

    Fig. 6. Grad-CAM heatmap overlays for the same test-set MRI scan (true label: meningioma), processed by (left) ResNet18, which correctly predicted meningioma, and (right) ResNet50, which misclassified this case as glioma. Warm colors (red/yellow) indicate higher gradient-weighted activation. This paired example illustrates that ResNet50s stronger aggregate meningioma recall (Section IV-E) does not mean it gets every meningioma case right ResNet18 correctly classified this particular scan while ResNet50 did not.

  5. LIMITATIONS

    While this study provides a structured, cost-aware com- parison of ResNet18 and ResNet50 for brain tumor MRI classification, several limitations should be acknowledged:

    • Single dataset: All experiments were conducted on a single publicly available brain MRI dataset. Performance and the relative ranking of the two architectures may not

    generalize to MRI data acquired from different scanners, imaging protocols, or patient populations.

    • Training epochs: Training was done for only 5 epochs in both models. It remains unknown whether the accuracy gap between ResNet18 and ResNet50 stays the same, decreases, or increases further with additional training epochs; ResNet50s validation accuracy dip at epoch 4 also suggests further training could change the picture.

    • No use of external validation: No independent valida- tion set from another institution or scanner was used to demonstrate the generalizability of the findings beyond the internal splits of the source dataset.

    • Statistical significance testing: A 5-seed statistical study was conducted (Section IV-F), but the results are mixed: the paired t-test indicates a significant difference (p = 0.036), while the Wilcoxon signed-rank test (p = 0.0625) and McNemars test on paired predictions (p = 0.551) do not. With only 5 seeds, statistical power is limited, and the multi-seed studys absolute accuracies (8688%) do not exactly match the single main run reported in Table II (88.989.1%), suggesting the two used different training configurations. A larger number of seeds under the exact main-run configuration would be needed to resolve this ambiguity conclusively.

    • Transfer-learning-only approach: Only the last classifi- cation layer was fine-tuned, while all other convolutional layers remained frozen with their pre-trained ImageNet weights. Fine-tuning of the residual layers was not con- sidered.

    • Incomplete measurement of computational costs: Pa- rameter counts, model size, FLOPs, inference latency, training time, and throughput were measured directly. Ad- ditional deployment-related metrics, such as peak mem- ory usage and energy consumption were outside the scope of this study and were therefore not evaluated.

  6. CONCLUSION

    This research deployed and assessed two convolutional neural architectures, ResNet18 and ResNet50, for categorizing brain MRI scans into four classes: glioma, meningioma, pitu- itary growth, and non-tumorous cases, using a systematically organized dataset with clearly separated training, validation, and testing partitions.

    Instead of the straightforward conclusion that ResNet50 is better, this paper contextualizes its results in terms of practical deployment. ResNet50 achieved a marginally better overall accuracy and F1-score in the main comparison (by about 0.2 percentage points) and a somewhat larger, though still modest, advantage across a 5-seed statistical study (about 1.1 percentage points on average). It showed a clearer benefit specifically for the harder-to-classify meningioma class; it did not outperform ResNet18 on glioma or pituitary in this run. This modest, class-specific improvement comes together with an increase in the number of parameters and, based on the metrics in Section IV-E, with substantially greater training time (roughly 2.5× per epoch) and lower training throughput.

    Statistical testing across seeds gave mixed resultssignificant by paired t-test but not by the Wilcoxon or McNemars tests (Section IV-F)so ResNet50s advantage, while directionally consistent, is not conclusively confirmed. As an alternative to ResNet50 in constrained environments such as point-of-care and low-resource healthcare, the equally efficient ResNet18 remains a defensible choice given this uncertainty. In contrast, when th application specifically prioritizes meningioma sensi- tivity and sufficient GPU resources are available, the additional cost of ResNet50 is more clearly justified.

    As a general recommendation, deployment decisions should be based on (i) the magnitude of the accuracy difference and its statistical significance, (ii) the computational cost of the model on the target hardware, and (iii) the clinical cost of specific errors (a false-negative glioma error is far more consequential than a false-positive no tumor error). Given that ResNet50 did not outperform ResNet18 on glioma in this run, there is no blanket reason to prefer the deeper network; the decision must be made not only on the basis of performance but also on the basis of cost and the specific class-level priorities of the deployment setting.

  7. FUTURE WORK

Building directly on the scope of this study, several concrete extensions are proposed:

    • Resolve the statistical ambiguity with more seeds under matched conditions: The 5-seed study (Sec- tion IV-F) gives conflicting significance results between the paired t-test and the Wilcoxon/McNemars tests. Repeating the seed sweep with a larger number of seeds (n 1520) under the exact same training configuration as the main comparison in Table II would improve sta- tistical power and resolve whether ResNet50s advantage is genuine or an artifact of the small sample.

    • Broader deployment profiling: Extend the evaluation by measuring additional deployment metrics such as peak memory usage and energy consumption on both CPU and GPU hardware.

    • Investigate the ResNet50 epoch-4 validation dip: The accuracy/loss curves (Section IV-E) show a pronounced dip in ResNet50s validation performance at epoch 4 that does not appear for ResNet18. Investigating whether this is caused by learning-rate sensitivity, batch-norm statis- tics, or a specific difficult batch could clarify whether ResNet50s training is inherently less stable in this setting or whether it is an artifact of this particular run.

    • Extended and planned training: Test the performance of ResNet18 and ResNet50 after increasing the amount of training beyond the current number of epochs, using learning-rate scheduling or early stopping based on vali- dation lossparticularly given ResNet50s validation dip at epoch 4 in this run.

    • External and multi-site validation: Evaluate both mod- els on independently sourced MRI datasets from different scanners and institutions to assess generalization beyond the single dataset used here.

  • Depth ablation / fine-tuning: Systematically compare transfer learning with a frozen backbone (as used in this study) against partial or complete fine-tuning of the resid- ual layers for both architectures, to determine whether the costbenefit ratio changes with greater adaptation of the model to the MRI domain.

  • Class-imbalance-aware training: Given the class-wise pattern observed here (ResNet50 helping meningioma but not glioma or pituitary), explore class-weighted loss functions or targeted data augmentation for underper- forming classes within the existing ResNet18/ResNet50 framework.

  • Quantitative Grad-CAM evaluation: Move beyond qualitative heatmap inspection toward quantitative local- ization metrics (e.g., overlap with radiologist-annotated tumor regions, where available) to more rigorously com- pare the interpretability of the two architectures.

REFERENCES

  1. K. He, X. Zhang, S. Ren, and J. Sun, Deep residual learning for image recognition, in Proc. IEEE Conf. Comput. Vis. Pattern Recognit. (CVPR), 2016, pp. 770778.

  2. M. Musthafa, T. R. Mahesh, V. Vinoth Kumar, and S. Guluwadi, Enhancing brain tumor detection in MRI images through explainable AI using Grad-CAM with ResNet50, BMC Med. Imag., vol. 24, no. 1, Art. no. 1292, 2024.

  3. H. Sahli, A. B. Slama, A. Zera¨i, S. Labidi, and M. Sayadi, ResNet- SVM: Fusion based glioblastoma tumor segmentation and classifica- tion, Comput. Biol. Med., vol. 152, p. 106438, 2023.

  4. G. Litjens, T. Kooi, B. E. Bejnordi, et al., A survey on deep learning in medical image analysis, Med. Image Anal., vol. 42, pp. 6088, 2017.

  5. S. Pereira, A. Pinto, V. Alves, and C. A. Silva, Brain tumor segmenta- tion using convolutional neural networks in MRI images, IEEE Trans. Med. Imag., vol. 35, no. 5, pp. 12401251, 2016.

  6. M. Hossain, et al., Brain tumor detection and classification in MRI using deep CNN models, J. Imag., vol. 7, no. 7, p. 123, 2021.

  7. R. Paul, et al., Deep learning-based brain tumor classification from MRI images, Comput. Methods Programs Biomed., vol. 187, p. 105113, 2020.

  8. M. A. Khan and R. Ullah, Improved CNN models for multi-class brain

    tumor classification, Comput. Biol. Med., vol. 133, p. 104167, 2021.

  9. E. Tjoa and C. Guan, A survey on explainable artificial intelligence (XAI): Toward medical XAI, IEEE Trans. Neural Netw. Learn. Syst., vol. 32, no. 11, pp. 47934813, 2020.

  10. R. R. Selvaraju, M. Cogswell, A. Das, et al., Grad-CAM: Visual explanations from deep networks via gradient-based localization, in Proc. IEEE Int. Conf. Comput. Vis. (ICCV), 2017, pp. 618626.